Postprandial response improvement via safety layer in closed-loop blood glucose controllers

Traditional type 1 diabetes therapies are prone to show poor glucose regulation especially in the postprandial period owing to both physiological and technological limitations. Although a closed-loop controller for glucose regulation has to be tuned to minimize the postprandial excursion and avoid late hypoglycemia, the intrinsic limitations of the problem lead to a trade-off between postprandial peak and late hypoglycemia risk. This paper reveals through an intensive in-silico study with multiple controller tuning combinations that a novel safety layer for glucose controllers, the so-called SAFE loop (Revert et al., 2013), not only reduces the hypoglycemia events but also allows reducing the postprandial glucose excursion, thus breaking the implicit trade-off present in single controllers. The SAFE outer loop monitors the estimated amount of insulin on board, and modifies the control action if it is close to a unique constraint which can be adjusted with clinical criteria. A very challenging test scenario is here implemented including the rate of blood glucose appearance from intakes of mixed meals, diurnal and day-to-day time-varying metabolic changes, inherent drawbacks in sensor and actuator, and other realistic conditions. The results show a significant reduction of hypoglycemia events when SAFE is added, regardless the closed-loop glucose controller, together with a potential postprandial response improvement

Research in this area is partially supported by the Spanish Ministry of Science and Innovation through grant DPI2010-20764-C02, the Government of Catalonia trough grant SGR00523, and the Argentinean Government (ANPCyT PICT-2011-0888 and 2012-0037, CONICET PIP112-2011-00361 and UNLP 1164). Fabian Leon-Vargas acknowledges the FI grants of the Government of Catalonia

Elsevier

Manager: Ministerio de Ciencia e Innovación (Espanya)
Generalitat de Catalunya. Agència de Gestió d’Ajuts Universitaris i de Recerca
Author: Garelli, Fabricio
León Vargas, Fabian Mauricio
Garelli, Fabricio
Battista, Hernán de
Vehí, Josep
Abstract: Traditional type 1 diabetes therapies are prone to show poor glucose regulation especially in the postprandial period owing to both physiological and technological limitations. Although a closed-loop controller for glucose regulation has to be tuned to minimize the postprandial excursion and avoid late hypoglycemia, the intrinsic limitations of the problem lead to a trade-off between postprandial peak and late hypoglycemia risk. This paper reveals through an intensive in-silico study with multiple controller tuning combinations that a novel safety layer for glucose controllers, the so-called SAFE loop (Revert et al., 2013), not only reduces the hypoglycemia events but also allows reducing the postprandial glucose excursion, thus breaking the implicit trade-off present in single controllers. The SAFE outer loop monitors the estimated amount of insulin on board, and modifies the control action if it is close to a unique constraint which can be adjusted with clinical criteria. A very challenging test scenario is here implemented including the rate of blood glucose appearance from intakes of mixed meals, diurnal and day-to-day time-varying metabolic changes, inherent drawbacks in sensor and actuator, and other realistic conditions. The results show a significant reduction of hypoglycemia events when SAFE is added, regardless the closed-loop glucose controller, together with a potential postprandial response improvement
Research in this area is partially supported by the Spanish Ministry of Science and Innovation through grant DPI2010-20764-C02, the Government of Catalonia trough grant SGR00523, and the Argentinean Government (ANPCyT PICT-2011-0888 and 2012-0037, CONICET PIP112-2011-00361 and UNLP 1164). Fabian Leon-Vargas acknowledges the FI grants of the Government of Catalonia
Document access: http://hdl.handle.net/2072/297006
Language: eng
Publisher: Elsevier
Rights: Tots els drets reservats
Subject: Glucèmia
Blood sugar
Hipoglucèmia
Hypoglycemia
Diabetis
Diabetes
Control, Teoria de
Control theory
Anàlisi de sistemes
System analysis
Control intel·ligent
Intelligent control systems
Control automàtic
Automatic control
Title: Postprandial response improvement via safety layer in closed-loop blood glucose controllers
Type: info:eu-repo/semantics/article
Repository: Recercat

Subjects

Authors